Factors Determining the Need For Bilevel Therapy in Obstructive Sleep Apnea Patients

J Turk Sleep Med 2023;10(1):78-83
1. University of Health Sciences Turkey, Samsun Training and Research Hospital, Clinic of Pulmonary Disease, Samsun, Turkey
No information available.
No information available
Received Date: 07.06.2022
Accepted Date: 19.08.2022
Publish Date: 14.02.2023
PDF
Cite
Share
Request

ABSTRACT

Objective:

Except for continuous positive airway pressure (CPAP), which is the first choice in the standard treatment of Obstructive Sleep Apnea Syndrome (OSAS), it is recommended to continue treatment with bilevel PAP (BPAP) in patients, who cannot tolerate constant pressure or no success was achieved with this treatment. However, there is a group of patients without complicated OSAS and do not have BPAP indication at the first hospitalization, but for which adequate titration cannot be performed with CPAP. The purpose of this study was to investigate which characteristics of these patients or which data in their polysomnography (PSG) may be indicative of BPAP need.

Materials and Methods:

Comorbid diseases [diabetes mellitus (DM), cardiac diseases, pulmonary diseases], body mass index, neck/core/hip circumference measurements the patients evaluated with total sleep time, apnea-hypopnea index (AHI), hourly obstructive/central apnea and hypopnea numbers, mean desaturation index (ODI), rapid eye movement (REM) sleep latency, REM time, AHI in REM and non-REM (REM/non-REM index), average overnight saturation (SaO2), lowest saturation value (min O2%), time when saturation is below 90% overnight (T90), a position dependency in PSG.

Results:

Presence of DM, hypertension and cardiac disease, elevation of neck/core/hip circumference measurements, ODI, REM index, T90 values and ODI/SaO2 ratios were found to be statistically significant in the BPAP group, elevation in min O2% and SaO2 levels were found to be statistically significant in the CPAP group (p<0.05). It was determined that the probability of BPAP increased with the presence of DM 0.214 times, the presence of heart disease 0.205 times, a one-unit increase in the REM index 1.018 times, and a one-unit increase in T90 1.030 times. The REM index and T90, which were found to be significant in the receiver operating characteristic analysis, were determined as 70.850 and 56.150 cut-off values, respectively.

Conclusion:

In this study; it was determined that the probability of CPAP being insufficient and switching to BPAP increases with the presence of DM and heart disease; and that T90 and REM index, and their cutoff values can be used for this purpose. It was also thought that regional adiposity may affect the type of PAP to be used.

Introduction

Obstructive sleep apnea syndrome (OSAS) is a disease setting that progresses with repetitive interruptions or slowdowns of the respiratory period, reduces sleep quality and causes frequent desaturation. According to the current guide of the American Academy of Sleep Medicine (AASM) (1); continuous positive airway pressure (CPAP) is the first choice of standard treatment in uncomplicated OSAS cases. It is recommended to continue treatment with bilevel PAP (BPAP) in patients in those no success could be achieved with CPAP treatment (continuing low saturation in the routine control at the 3rd month under treatment) or in patients who cannot tolerate constant pressure. BPAP as a first choice, can be used in the presence of a disease that causes difficulty in exhaling against high pressure or alveolar hypoventilation (such as obesity hypoventilation syndrome) (1,2).

In some randomized studies performed, BPAP treatment was similar to CPAP in terms of effectiveness, adaptation, and some endpoints (3,4). In studies in which treatments started with BPAP instead of CPAP were evaluated in order to increase adaptation with uncomplicated patients, no results in favor of BPAP could be achieved (5,6). Among the recommendations for manual titration in the guideline of the AASM, it is stated that switching to BPAP could be performed in cases where CPAP could not be tolerated or is insufficient. However, there is a group of patients who do not have complicated OSAS and do not have BPAP indication at the first hospitalization, but for which adequate titration cannot be performed with CPAP. Repeated hospitalizations of this group and the prolonged access to definitive treatment create problems both in terms of patient health and comfort and the faster and more economical functioning of the health system. There is limited data on which characteristics of such patients or which data in their polysomnography (PSG) may be indicative of BPAP need (7,8). 

In this study, it was planned to investigate which characteristics of the patient group who did not benefit from CPAP treatment and were switched to BPAP treatment, and which variables in PSGs could be similar and directing. By this way, pre-recognition of this patient group, providing an advantage in terms of recurrent hospitalizations, cost and time loss were determined as a secondary target.

Materials and Methods

Study Design and Participants

This study was designed as an observational study in which patients hospitalized in the sleep disorders unit of Samsun Training and Research Hospital were retrospectively examined. Patients over the age of 18 who had an apnea-hypopnea index (AHI) >15/h as a result of all-night PSG and were diagnosed with sleep apnea and were deemed suitable for PAP treatment were randomly included in the study.

After the titration study, the data of the patients for whom CPAP treatment was approved and the patients whose CPAP device was insufficient and switched to BPAP titration, were evaluated. Patients with known chronic respiratory failure or sleep-related hypoventilation/hypoxemic syndromes, who started treatment with BPAP or BPAP/ST titration, and who meet the criteria of obesity hypoventilation syndrome [combination of obesity (body mass index (BMI) ≥30 kg·m-2), sleep disordered breathing after ruling out other disorders that may cause alveolar hypoventilation and confirmed daytime hypercapnia] were not included in the study.

Study Data

Ethical approval was taken from Samsun University Clinical Trials Ethic Committee (protocol no: SUKAEK-2022/1/5, date no: 18/05/2022).

Demographic data and first hospitalization all-night PSG results of the patients who met the criteria and were hospitalized for titration between January 2019 and March 2022 in the sleep laboratory were examined.

Standard 16-channel PSG (Embla; Medcare Flaga, Reykjavik, Iceland) was applied to all patients. The system consists of four electroencephalography channels, two electrooculography channels, tibial/submental electromyography, and electrocardiography, as well as monitoring of oronasal airflow, thoracic movements, abdominal movements, average overnight saturation (SaO2), and body position. PSGs were scored in accordance with the 2013 AASM guidelines. Manual titration attended to moderate or severe OSAS patients. PAP titration was switched to BPAP, if a patient was uncomfortable or intolerant of CPAP or CPAP was ineffective at controlling respiratory events.

Gender, age, comorbid diseases [diabetes mellitus (DM), hypertension (HT), cardiac diseases, pulmonary diseases], BMI, neck/core/hip circumference lengths of the patients were recorded. In PSG; total sleep time, AHI, hourly obstructive/central apnea and hypopnea numbers, mean desaturation index (ODI), rapid eye movement (REM) sleep latency, REM time, REM and non-REM index, SaO2, lowest saturation value (min O2%), time when saturation is below 90% overnight (T90), whether it is position dependent were evaluated.

Statistical Analysis

In the statistical analysis phase of the study, frequency analysis, descriptive statistics, chi-square relationship tests, mean comparison tests, multiple logistic regression analysis and receiver operating characteristic (ROC) curves were shown. The conformity of numerical measurements to normal distribution was examined with the Shapiro-Wilk test and the independent samples were examined with t-test. Sensitivity, 1-selectivity (solid-phase extraction), positive class prediction rate, cut-off and area under the curve values are provided along with the graph of ROC curves. The entire application was performed with the R-Project program (R Core Team, 2020).

Results

A total of 111 patients, 31 (27.9%) of whom BPAP and 80 (72.1%) of whom CPAP titration was applied, were included in the study. The gender distribution and comorbidity status of the patients are shown in Table 1, and the mean values of the numerical data are shown in Table 2.

Table 1
Table 2

When the relationship between gender, positional OSAS, pulmonary disease, DM, HT, presence of coronary artery disease (CAD) or heart failure (HF) and receiving CPAP and BPAP treatment of the patients was compared (Table 3); the presence of DM, HT and cardiac disease was found to be statistically significantly higher in patients in the BPAP group (p<0.05).

Table 3

The comparison of the mean values between the variables is provided in Table 4. Accordingly, neck/core/hip circumference measurements, ODI, REM index, T90 and ODI/SaO2 ratio were found to be statistically significantly higher in the BPAP group and, min O2% and SaO2 were found to be statistically significantly higher in the CPAP group (p<0.05).

Table 4

Multiple regression analysis is provided in Table 5. Accordingly, the suitability of BPAP therapy to the patient is increased 0.214 times in the presence of DM and 0.205 times in the presence of CAD. Moreover, it was determined that a one-unit increase in REM index increased the probability of BPAP 1,018 times, and a one-unit increase in T90 increased the probability of BPAP 1.030 times.

Table 5

REM index and T90% were found to be statistically significant in the ROC evaluation performed on the variables that were significant according to the result of multiple regression analysis (Figures 1 and 2). 70.850 for REM index and 56.150 for T90 were determined as cut-off values.

Figure 1
Figure 2

Discussion

There is a small group of patients with uncomplicated OSAS who did not benefit from CPAP titration and were switched to BPAP. Due to the lack of guiding data in terms of pre-identifying these patients and initiating appropriate treatment quickly, this study was planned and it was concluded that some variables could be used for this purpose. Accordingly, especially after multiple regression analysis, the presence of DM and cardiac disease increased the possibility of transition to BPAP, and T90 and REM index were determined as determinant variables for BPAP, cut-off values were obtained for T90 and REM index.

BPAP is an approach that increases the patient’s compliance with PAP treatment by applying different pressures during inspiration and expiration. In studies investigating the reasons for switching to BPAP in OSAS patients, the main reason was often intolerance to CPAP (7,9,10). In the study performed by Schwartz et al. (10), it was reported that treatment with BPAP can be started in OSAS patients with neurological diseases in whom sufficient muscle tone cannot be achieved. However, in the guide published by AASM in 2019, it has been suggested that BPAP can be used as an initial treatment or as a subsequent treatment to CPAP on the same night, with the clinician’s decision, in suitable patients (11). However, there is a lack of data on which patient should be switched to BPAP using which criteria.

In this study, in which we investigated BPAP indications, when comorbidities were evaluated with the chi-square relationship test, the presence of DM, HT and heart disease was found to be significantly higher in terms of BPAP treatment (p<0.05). In the multiple regression analysis including other variables, the probability of needing BPAP was found to be 0.214 times higher in the presence of DM and 0.205 times higher in the presence of heart disease. HF is a disease associated with OSAS at a rate of approximately 33%. While CPAP therapy increases the left heart ejection fraction by approximately 5% in patients with systolic dysfunction, no data are available as the use of PAP is contraindicated in the setting of diastolic failure. CPAP therapy reduces major adverse cardiac or cerebrovascular events in CAD patients with OSAS (12). There is no clarity on the effect of CPAP on short- and long-term blood glucose control and diabetes-related complications in DM patients due to inconsistent results (13). If there is no contraindicated condition in this group of patients, randomized controlled studies in which BPAP treatment will be applied after the clinician’s evaluation will provide us more data on this subject.

Among the risk factors for OSAS, obesity is the most important one; however, while 70% of OSAS patients are obese, 40% of obese patients have OSAS (14,15), and the reason for this is still unclear (16). It was reported that high BMI worsens the response to treatment due to the patient’s general body performance rather than directly affecting CPAP therapy (17). In the study performed by Shah et al. (7), BMI was found to be high in the BPAP group, but this did not make a statistical difference, and it was reported that it was not a factor for CPAP failure. In our study, a significant difference was observed between the groups not in terms of BMI, but in terms of neck/core/hip circumference measurements. We think that the effect of regional adiposity on the absence of OSAS in every patient with high BMI and the lack of similar benefit from CPAP, should be investigated with larger case series.

Slouka et al. (17) reported that AHI, ODI, T90, SpO2 and BMI are significant variables for CPAP failure, but after the ROC analysis, they reported that any one or combinations of those variables were insufficient in modeling and clinical prediction. In another study in which age, AHI, COPD, BMI and min O2% were determined as independent variables for BPAP being the first treatment option (18), there was no significant difference between those who received CPAP treatment in terms of adherence to treatment, AHI under treatment, or symptoms. Schwartz et al. (10) reported that advanced age, high BMI, COPD, increased CO2 level, OSAS severity, and min O2% values could be determinant indicators not only for BPAP use but also for long-term adherence to treatment.

In our study, in terms of PSG variables, significant results were obtained for ODI, REM index, T90, min O2%, SaO2 values and ODI/SaO2 ratios. ODI, T90, min O2% and SaO2 were prominent variables in previous studies in terms of CPAP failure or BPAP indication (10,17,18). In our study, the ODI/SaO2 ratio was used for the first time in terms of BPAP estimation, and it was determined that as this ratio increased, the need for BPAP increased, but a statistically significant numerical ratio could not be obtained.

In the ROC analysis, one of the most determinant variables in terms of the statistical model was T90. A 1-unit increase in this variable increases the probability of BPAP 1,030 times. In another study in which T90 was used for this purpose (7), it was stated that with a 5% increase, the probability of BPAP increases by 28%. We also predict that the value of 56.15 for T90 is considered cut-off, and values above this can be used for BPAP indication.

A commonly used definition for REM-dependent OSAS (REM-OSAS) is AHIREM >5 and AHINREM <5 in the presence of at least 30 minutes of REM sleep (19). However, its prevalence is not clear due to the lack of a standard definition (13). In a study conducted with this patient group (20), it was reported that T90 and min O2% should be included in the definition of REM-OSAS. It was stated in the limited number of studies that they could benefit from CPAP treatment (10,21), but there are no studies conducted with other PAP treatments in this group. The REM index, which indicates the AHI during REM sleep, was another determinant variable in our study. An increase of 1 unit increased the probability of BPAP 1.018 times, and values above 70.85 were found to be significant in terms of BPAP. In our study, which was carried out in accordance with the definition above, there was no patient diagnosed with REM-OSAS.

Not using different criteria for the definition of REM-OSAS was one of the limitations of our study. Because the number of patients diagnosed could have changed, and perhaps data on the use of BPAP in this group would have been available for the first time. In addition, the limited number of BPAP patients and the fact that the study was conducted with data from a single center can be listed as other limitations.

Conclusion

Recurrent hospitalizations in centers where the number of patients who need PSG and the waiting time for the test are high, make it difficult for the patient to access treatment, cause obstruction in the health system and increase costs. The inadequacy of CPAP and the necessity of switching to BPAP is the most important reason for this. In this study, which was designed to predict this group of patients, we determined that especially T90, REM index and the cut-off values obtained for these can be used for BPAP estimation. In addition, we think that there is a need for studies to be performed with more patients to examine the effect of regional adiposity on the type of PAP to be used.

Ethics

Ethics Committee Approval: Ethical approval was taken from Samsun University Clinical Trials Ethic Committee (protocol no: SUKAEK-2022/1/5, date no: 18/05/2022).

Informed Consent: Retrospective study.

Peer-review: Internally and externally peer-reviewed.

Financial Disclosure: The author declared that this study received no financial support.

References

1
Patil SP, Ayappa IA, Caples SM, Kimoff RJ, Patel SR, Harrod CG. Treatment of Adult Obstructive Sleep Apnea with Positive Airway Pressure: an American Academy of Sleep Medicine systematic Review, Meta-Analysis, and GRADE Assessment. J Clin Sleep Med 2019;15:301-34.
2
Freedman N. Treatment of Obstructive Sleep Apnea: Choosing the Best Positive Airway Pressure Device. Sleep Med Clin 2017;12:529-42.
3
Powell ED, Gay PC, Ojile JM, Litinski M, Malhotra A. A pilot study assessing adherence to auto-bilevel following a poor initial encounter with CPAP. J Clin Sleep Med 2012;8:43-47.
4
Gulati A, Oscroft N, Chadwick R, Ali M, Smith I. The impact of changing people with sleep apnea using CPAP less than 4 h per night to a Bi-level device. Respir Med 2015;109:778-83.
5
Gay P, Weaver T, Loube D, Iber C; Positive Airway Pressure Task Force; Standards of Practice Committee; American Academy of Sleep Medicine. Evaluation of positive airway pressure treatment for sleep related breathing disorders in adults. Sleep 2006;29:381-401.
6
Reeves-Hoché MK, Hudgel DW, Meck R, Witteman R, Ross A, Zwillich CW. Continuous versus bilevel positive airway pressure for obstructive sleep apnea. Am J Respir Crit Care Med 1995;151:443-9.
7
Shah S, Smotherman C, Louis M. Determinants of bilevel therapy in the management of obstructive sleep apnea. Sleep Breath 2021;25:1181-6.
8
Setareh J, Malekmohammad M, Amini M, Khoundabi B, Naghan PA. Useful variables during wake and sleep for prediction of positive airway pressure in obstructive sleep apnea titration. Sleep Science Practice 2019;3. doi: 10.1186/s41606-019-0033-7
9
Goyal A, Joshi A, Mitra A, Khurana A , Chaudhary P. Predictive factors for CPAP failure in obstructive sleep apnea patients. Lung India 2021;38:540-4.
10
Schwartz SW, Rosas J, Iannacone MR, Foulis PR, Anderson WM. Correlates of a prescription for Bilevel positive airway pressure for treatment of obstructive sleep apnea among veterans. J Clin Sleep Med 2013;9:327-35.
11
Patil SP, Ayappa IA, Caples SM, Kimoff RJ, Patel SR, Harrod CG. Treatment of Adult Obstructive Sleep Apnea with Positive Airway Pressure: An American Academy of Sleep Medicine Clinical Practice Guideline. J Clin Sleep Med 2019;15:335-43.
12
Capodanno D, Milazzo G, Cumbo M, Marchese A, Salemi A, Quartarone L, Benvenuto E, Galseran C, Distefano SM, Tamburino C. Positive airway pressure in patients with coronary artery disease and obstructive sleep apnea syndrome. J Cardiovasc Med (Hagerstown) 2014;15:402-6.
13
Freedman N. Treatment of Obstructive Sleep Apnea Choosing the Best Positive Airway Pressure Device. Sleep Med Clin 2017;12:529-42.
14
Wolk R, Shamsuzzaman AS, Somers VK. Obesity, sleep apnea, and hypertension. Hypertension 2003;42:1067-74.
15
Vgontzas AN, Tan TL, Bixler EO, Martin LF, Shubert D, Kales A. Sleep apnea and sleep disruption in obese patients. Arch Intern Med 1994;154:1705-11.
16
Mokhlesi B, Tulaimat A, Faibussowitsch I, Wang Y, Evans AT. Obesity hypoventilation syndrome: prevalence and predictors in patients with obstructive sleep apnea. Sleep Breath 2007;11:117-24.
17
Slouka D, Honnerova M, Hosek P, Matas A, Slama K, Landsmanova J, Kucera R. Risk factors for failure of continuous positive airway pressure treatment in patients with ostructive sleep apnoea. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub 2018;162:134-8.
18
Mansukhani MP, Kolla BP, Olson EJ, Ramar K, Morgenthaler TI. Bilevel positive airway pressure for obstructive sleep apnea. Expert Rev Med Devices 2014;11:283-94.
19
Mokhlesi B, Punjabi NM. “REM-related” obstructive sleep apnea: an epiphenomenon or a clinically important entity? Sleep 2012;35:5-7.
20
Oweidat KA, AlRyalat SA, Al-Essa M, Obeidat N. Comparing REM- and NREM-Related Obstructive Sleep Apnea in Jordan: A Cross-Sectional Study. Can Respir J 2018;2018:9270329. doi: 10.1155/2018/9270329
21
Su CS, Liu KT, Panjapornpon K, Andrews N, Foldvary-Schaefer N. Functional outcomes in patients with REM-related obstructive sleep apnea treated with positive airway pressure therapy. J Clin Sleep Med 2012;8:243-7.